EPHIN Trends Post-Failure-mode-A On

EPHIN E150 Coincidence Rate

The E150 rate shows an increase on approaching the Earth's radiation zone, likely due to electrons. Additionally there is a gradual upward trend in the rate as the sensor temperature increases. The rate increase with temperature is due to a decrease in detector C sensitivity (see Figure 8), resulting in a mis-assignment of higher energy particles into this coincidence channel.

E150 rate vs time
Figure 1: EPHIN E150 coincidence channel rate vs time for the time outside the radiation zone of the orbits following the reconfiguration of EPHIN with failure-mode detector A on and the detector B ring segments disabled. The horizontal red line is the RADMON trigger threshold, 800000 cts/sample.

E150 rate vs EPHIN sensor temperature
Figure 2: EPHIN E150 coincidence channel rate vs EPHIN sensor temperature for times as selected in figure 1. The sensor temperature values have a small, uniformly-distributed random valued added for display purposes. The red curve is the median rate as a function of temperature.


"E1300" Rate

The E1300 rates show drops associated with higher EPHIN sensor temperatures. The decrease in rate with higher temperature is due to a decrease in detector C sensitivity as the temperture increases (see Figure 8).

E1300 rate vs time
Figure 3: EPHIN E1300 coincidence channel rate vs time for times as selected in figure 1. The horizontal red line is the RADMON trigger threshold 1000 cts/sample

E1300 rate vs EPHIN sensor temperature
Figure 4: EPHIN E1300 coincidence channel rate vs EPHIN sensor temperature for times as selected in figure 1. The red curve is the median rate as a function of temperature.


Detector B0 (Center Segment)

The detector B0 rate increases as the Earth's radiation zone is approached, similar to the increases in the E150 rate. The rate does not appear to be affected by sensor temperature over the observed range.
Detector B0 rate vs time
Figure 5: EPHIN detector B0 rate vs time for times as selected in figure 1.

Detector B0 rate vs EPHIN sensor temperature
Figure 6: EPHIN detector B0 rate vs EPHIN sensor temperature for times as selected in figure 1.


Detector C

The detector C rate decreases with increasing sensor temperature above roughly 43 degrees C. This behavior is the same as observed prior to setting detector A failure-mode on. The decrease in rate implies a decreased sensitivity to particles, possibly due to a lower voltage across the detector.
Detector C rate vs time
Figure 7: EPHIN detector C rate vs time for times as selected in figure 1.

Detector C rate vs EPHIN sensor temperature
Figure 8: EPHIN detector C rate vs EPHIN sensor temperature for times as selected in figure 1.


Detector D

As seen in the past, the detector D rate does not appear to be affected by sensor temperature over the observed range. Over the period examined the temperature has not been as high as during the Normal-Sun-Mode episodes and did not reach the temperature where the rate has been observed to drop.
Detector D rate vs time
Figure 9: EPHIN detector D rate vs time for times as selected in figure 1.

Detector D rate vs EPHIN sensor temperature
Figure 10: EPHIN detector D rate vs EPHIN sensor temperature for times as selected in figure 1.


Detector E

Similar to detector D, the detector E rate does not appear to be affected by sensor temperature over the observed range. There was a brief episode of noisy behavior in the rate near the start of the second orbit similar to noise observed in the past. This noise is similar to that observed in the same detector on the SOHO EPHIN and has been attributed by the EPHIN IPI team to mounting stresses.
Detector E rate vs time
Figure 11: EPHIN detector E rate vs time for times as selected in figure 1.

Detector E rate vs EPHIN sensor temperature
Figure 12: EPHIN detector E rate vs EPHIN sensor temperature for times as selected in figure 1.


Detector F

Similar to detector D, the detector F rate does not appear to be affected by sensor temperature over the observed range.
Detector F rate vs time
Figure 13: EPHIN detector F rate vs time for times as selected in figure 1.

Detector F rate vs EPHIN sensor temperature
Figure 14: EPHIN detector F rate vs EPHIN sensor temperature for times as selected in figure 1.


Last modified: Thu Aug 20 10:51:09 EDT 2009


Dr. Michael Juda
Harvard-Smithsonian Center for Astrophysics
60 Garden Street, Mail Stop 70
Cambridge, MA 02138, USA
Ph.: (617) 495-7062
Fax: (617) 495-7356
E-mail: mjuda@cfa.harvard.edu