AXAF Constraints, Restrictions, & Limitations

Number: HRC-C-006

Type: Constraint

Original Date: 1/31/97


Subsystem: HRC

Subject: HRC High Voltage and Detector Selection

Source: Informal CARD inputs, e-mail, M. Juda, 1/28/97


CRL Statement:

HRC high voltages must not be applied to a detector, i.e., Imaging or Spectroscopy, unless that detector has been selected by the operating electronics.


CRL Rationale/Justification:

HRC sensitivity to the application of high voltage to detectors is derived from the HRC design.


CRL Impact:

Inappropriate application of HRC high voltages to a detector will damage the detector grids.


Recommended Action:

Incorporate into the Systems and Operations Procedures, with ASC providing the inputs.