About Chandra Archive Proposer Instruments & Calibration Newsletters Data Analysis HelpDesk Calibration Database NASA Archives & Centers Chandra Science Links

Skip the navigation links
Last modified: 4 September 2008

URL: http://cxc.harvard.edu/ciao3.4/why/acistgain.html
Hardcopy (PDF): A4 | Letter

ACIS Time-dependent Gain


Summary

The tool acis_process_events includes a time-dependent gain (TGAIN) adjustment. This adjustment is necessary because the "effective gains" of the detectors are drifting with time as the result of an increasing CTI.

There is TGAIN calibration data for the entire ACIS detector (chips I0-3 and S0-5) at the -120 C focal plane temperature.

There is also TGAIN calibration data for the ACIS-S1 and S3 back-illuminated chips at the -110 C focal plane temperature.

If the data have gone through Reprocessing III, a TGAIN correction has been applied. Repro III started with DS 7.6.7. In some cases, there may be a new TGAIN available, requiring the user to reprocess the data.

The observation date and focal plane temperature are recorded in the header of the event file:

unix% dmkeypar acis_evt2.fits DATE-OBS echo+
2008-01-19T20:41:39

unix% dmkeypar acis_evt2.fits FP_TEMP echo+
153.60722351

What Analyses are Affected?

Users interested in CCD spectroscopy should apply the gain adjustment. Grating spectroscopy benefits as well, in the form of improved order sorting. It is not necessary to apply this adjustment if you are only doing timing or imaging analysis, but it will not have a negative effect, either.

Determine the Size of the Adjustment

The show_tgain_corr script displays the size of the time-dependent gain correction - i.e. how much the energy or PHA value of an event changes - for a given source location. The most recent version of the show_tgain_corr script is v1.6 (16 April 2007):

unix% grep Id `which show_tgain_corr`
% $Id: show_tgain_corr,v 1.6 2007/04/16 16:39:17 egalle Exp $

show_tgain_corr calls a helper script named show_tgain_corr.sl. The most recent version of that script is v1.7 (13 Sept 2007):

unix% grep Id $ASCDS_CONTRIB/share/slsh/local-packages/show_tgain_corr.sl
% $Id: show_tgain_corr.sl,v 1.7 2007/09/13 10:44:01 egalle Exp $

Note that $ASCDS_CONTRIB/share/slsh/local-packages/ is the default path in the standard CIAO scripts installation. If you do not have the scripts installed or need to update to a newer version, please refer to the Scripts page. To run show_tgain_corr, supply an ACIS chip number and the desired location in chip coordinates, e.g.

unix% show_tgain_corr 3 962 964

unix% show_tgain_corr 7 221 532

These command create a plot of the adjustment in energy for an on-axis source observed with chip I3 and chip S3, respectively. The plot compares the newest TGAIN files to the previous default versions.

The script includes options to plot in PHA and to store the output as a PS or GIF file; the help file for the script (ahelp show_tgain_corr) has more information.

Improvements in Calibration

Hardcopy (PDF): A4 | Letter
Last modified: 4 September 2008


The Chandra X-Ray Center (CXC) is operated for NASA by the Smithsonian Astrophysical Observatory.
60 Garden Street, Cambridge, MA 02138 USA.    Email: cxcweb@head.cfa.harvard.edu
Smithsonian Institution, Copyright © 1998-2004. All rights reserved.